Aalborg DFC Digital Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Transmissão Aalborg DFC Digital. Aalborg DFC Digital User Manual [fr] Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 0
OPERATING MANUAL
DFC MASS FLOW
CONTROLLER
Technical Data Sheet No. TD9805M Rev. O
Date of Issue: February 2012
Vista de página 0
1 2 3 4 5 6 ... 43 44

Resumo do Conteúdo

Página 1 - OPERATING MANUAL

OPERATING MANUALDFC MASS FLOW CONTROLLERTechnical Data Sheet No. TD9805M Rev. ODate of Issue: February 2012

Página 2 - TABLE OF CONTENTS

4.1 DFC 26/36/46 Mass Flow ControllersACCURACY: +1% of full scale, including linearity for gas temperatures ranging from59 FF to 77 FF (15 FC to 25 F

Página 3

Aalborg7 makes no expressed or implied guarantees of corrosion resistance ofmass flow meters as pertains to different flow media reacting with compone

Página 4

FLOW RANGESTABLE I DFC 26 LOW FLOW MASS FLOW CONTROLLERS*TABLE II DFC 36 MEDIUM FLOW MASS FLOW CONTROLLERS*TABLE III DFC 46 HIGH FLOW MASS FLOW CONTRO

Página 5

TABLE IV PRESSURE DROPSTABLE V APPROXIMATE WEIGHTS10MODELFLOW RATE[std liters/min]MAXIMUM PRESSURE DROP[mm H2O] [psid][mbar]DFC 26up to 10720 1.06 75D

Página 6

5. OPERATING INSTRUCTIONS5.1 Preparation and Warm UpIt is assumed that the Mass Flow Controller or Controller has been correctlyinstalled and th

Página 7

5.4 Set Point Reference SignalDFC flow controllers have a built-in solenoid valve and allow the user to set theflow to any desired flow rate within

Página 8

5.7 Analog Interface ConfigurationThe DFC can be configured for the desired range and scaling by selection ofanalog board (see APPENDIX 1 on page 21

Página 9

6.2 Flow Path CleaningBefore attempting any disassembly of the unit for cleaning, try inspecting the flowpaths by looking into the inlet and outlet

Página 10

Inspect the flow path inside the transducer for any visible signs of contaminants.If necessary, flush the flow path through with alcohol. Thoroughly d

Página 11

priate relative correction factor should be recalculated see section (9).It is standard practice to calibrate Mass Flow Meters/Controllers with dry ni

Página 12 - FLOW RANGES

TABLE OF CONTENTS1. UNPACKING THE DFC MASS FLOW CONTROLLER...1.1 Inspect Package for External Damage...

Página 13 - TABLE V APPROXIMATE WEIGHTS

8. TROUBLESHOOTING8.1 Common ConditionsYour Mass Flow Controller was thoroughly checked at numerous quality controlpoints during and after manufac

Página 14

8.3 Troubleshooting Guide18Remedycheck connection of power supplydisconnect DFC transducer from powersupply; remove the shorting conditionor check pol

Página 15

19Indicationfull scale output at "no flow" condition orwith valveclosedcalibration offDFC valvedoes not workin open positionDFC valvedoes no

Página 16

20QO = Qa = Qr x K = 1000 X 0.9926 = 992.6 sccmwhere K = relative K factor to reference gas (oxygen to nitrogen)1d X Cpwhere d = gas dens

Página 17

APPENDIX 1COMPONENTS DIAGRAMS21DFC Digital PC Board (Primary Side)

Página 18

APPENDIX 1(CONTINUED)22DFC Digital PC Board(Secondary Side)

Página 19

APPENDIX 1(CONTINUED)23DFC Analog PC Board(Primary Side)

Página 20

APPENDIX 1(CONTINUED)24DFC Analog PC Board(Secondary Side)

Página 21 - 8.3 Troubleshooting Guide

25APPENDIX 2GAS FACTOR TABLES (“K” FACTORS)CAUTION: K-Factors at best are only an approximation. K factors should not be used in applications that

Página 22

26ACTUAL GASK FACTORRelative to N2Cp[Cal/g]Density[g/I]Deuterium D21.00 1.722 1.799Diborane B2H6.4357 .508 1.235Dibromodifluoromethane CBr2F2.1947 .15

Página 23

APPENDIX 1 COMPONENT DIAGRAM...APPENDIX 2 GAS FACTOR TABLE ("K" FACTORS)...

Página 24 - DFC Digital PC Board

27ACTUAL GASK FACTORRelative to N2Cp[Cal/g]Density[g/I]Hydrogen Bromide HBr1.000 .0861 3.610Hydrogen Chloride HCl1.000 .1912 1.627Hydrogen Cyanide HCN

Página 25 - (Secondary Side)

28ACTUAL GASK FACTORRelative to N2Cp[Cal/g]Density[g/I]Phosphorous Oxychloride POCl3.36 .1324 6.843Phosphorous Pentafluoride PH5.3021 .1610 5.620Phosp

Página 26 - DFC Analog PC Board

29APPENDIX 3DIMENSIONAL DRAWINGSDFC 26 Mass Flow ControllerNOTES: Aalborg7 reserves the right to change designs and dimensions at itssole discretion

Página 27

30DFC 36/46 Mass Flow ControllerNOTES: Aalborg7 reserves the right to change designs and dimensions at itssole discretion at any time without notice.

Página 28

APPENDIX 4SENDING COMMANDS TO THE DFCRS485The standard DFC comes with an RS485 interface. The protocol described below allows for the unit using eith

Página 34

1. UNPACKING THE DFC MASS FLOW CONTROLLER1.1 Inspect Package for External DamageYour DFC Mass Flow Controller was carefully packed in a sturdy

Página 35

37APPENDIX 5CALIBRATION TABLE: GAS INDEPENDENT VARIABLESINDEXNAMEDATATYPENOTES0 BlankSDPROC char[10] Do not modify. For internal use only.1 SerialNum

Página 36

38INDEX NAMEDATATYPENOTES29 Kgain[2] float30 Kgain[3] float31 Kgain[4] float32 Kgain[5] float33 Reserved float34 Reserved float35 Reserved float36 Res

Página 37

CALIBRATION TABLE: GAS DEPENDENT VARIABLES39INDEX NAMEDATATYPENOTES100 GasIdentifer char[27]101 FullScaleRange float102 StdTemp float103 StdPressure f

Página 38

40INDEX NAMEDATATYPENOTES129 SensorTbl[8][Sensor Value] uint130 SensorTbl[8][Flow] float131 SensorTbl[9][Sensor Value] uint132 SensorTbl[9][Flow] floa

Página 39

APPENDIX 6NOTE: Follow Return Procedures In Section 1.3.41WARRANTYAalborg7 Mass Flow Systems are warranted against parts and workmanshipfor a period

Página 40 - APPENDIX 5

Attitude sensitivity of the Mass Flow Controller is +15F. This means that the gas flowpath of the Flow Controller must be horizontal within those stat

Página 41

RS-485 option:The RS485 converter/adapter has to be configured for: multidrop, 2 wire, half duplexmode. The transmitter circuit has to be enabled by T

Página 42

4FIGURE b-1, WIRING DIAGRAM FOR DFC TRANSDUCERS.

Página 43

PIN FUNCTION1 +15 VDC Power Supply2 0-5 VDC Flow Signal (4-20mA Option)3 0-5 VDC Set Point Input (4-20mA Option) 4 Force Valve Open Control5 Forc

Página 44 -  CAUTION:

3. PRINCIPLE OF OPERATIONThe stream of gas entering the Mass Flow transducer is split by shunting a smallportion of the flow through a capillary st

Comentários a estes Manuais

Sem comentários