UART ERROR CODES
1 Not Supported Command or Back Door is not enabled.
2 Wrong # of Arguments.
3 Address is Out of Range (MR or MW commands).
4 Wrong # of the characters in the Argument.
5 Attempt to Alter Write Protected Area in the EEPROM.
6 Proper Command or Argument is not found.
7 Wrong value of the Argument.
8 Reserved.
9 Manufacture specific info EE KEY (wrong key or key is disabled).
DIAGNOSTIC AND SYSTEM EVENTS CODES AND BIT POSITION
Code Event Description Bit position
0 CPU Temp. High 0x0001
1 High Flow Alarm 0x0002
2 Low Flow Alarm 0x0004
3 Range between H-L 0x0008
4 Tot#1 > Limit 0x0010
5 Tot#2 > Limit 0x0020
6 OptPulse Queue 0x0040
7 Flow OverLimit 0x0080
8 Vcc OutOfRange 0x0100
9 SerComm. ERROR 0x0200
A EEPROM ERROR 0x0400
B Power on Event 0x0800
C Password Event 0x1000
D Fatal Error 0x2000
INTERNAL K-FACTOR TABLE LIST
Index Gas K-Factor
1 Ar 1.4573
2 AsH3 0.6735
3 BF3 0.5082
4 Br2 0.8083
5 C2H2 0.5829
6 C2N2 0.6100
7 CH4 0.7175
8 Cl2 0.8600
9 CO2 0.7382
10 COF2 0.5428
11 COS 0.6606
12 CS2 0.6026
13 F2 0.9784
14 H2 1.0106
15 He 1.4540
16 N2O 0.7128
17 NH3 0.7310
18 NE 1.4600
19 NO 0.9900
20 O2 0.9926
21 SO2 0.6900
22 Xe 1.4400