Aalborg Vortex Insertion Manual do Utilizador Página 88

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 87
83
COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNTAX
COMMAND
ARGUMENT 1
ARGUMENT 2
ARGUMENT 3 ARGUMENT 4 RESPONSE
Temperature
Alarms**
Sets / reads the status of the
temperature alarms.
Note: High and Low limits have
to be entered in the °C. High
alarm value has to be more
than Low alarm value.
Alarm conditions:
Temp. > High Limit = H
Temp. < Low Limit = L
Low < Temp. < High = N
9 TA H
(high temp. limit)
<Value>
(-20.1-260 °C)
TAH:<Value [°C]>
L
(low temp. limit)
<Value>
(-20.1-260 °C)
TAL:<Value [°C]>
A
(action delay in
seconds)
<Value> (0-3600
sec.)
TAA:<Value [Sec]>
E
(enable alarm)
TA:E
D
(disable alarm)*
TA:D
R (read current
status)
TAR:N (no alarm)
TAR:H (high alarm)
TAR: L (low alarm)
S (Read current
settings)
TAS:M,L,H,D,B where:
M – mode (E/D)
L – Low settings (°C)
H – High settings (°C)
D – Action Delay (sec)
B – Latch mode (0-1)
B Block (Latch)
mode
<Value>
(0-disabled*)
(1-enabl’d )
TAB:<Value> where:
Value = 0 – 1
Vista de página 87